Boekel Scientific Digital Dry Bath Incubators User Manual

Page 2

Advertising
background image

N2400048, REV 1 06/11

SPECIFICATIONS AND PERFORMANCE

Dimensions:

Accessory Heating

Accommodates

2 Block Model 113002

8” w x 9.8” d x 3.5” h

Blocks

113002-2

(20.3 x 24.9 x 8.6 cm)

110006

30 tubes up to 6mm

4 Block Model 113004

10.25” w x 11.25” d x 3.5” h

110010

20 tubes up to 10mm

113004-2

(26.0 x 28.6 x 8.9 cm)

110013

20 tubes up to 13mm

Net Weight:

110016

6 tubes up to 25mm

2 Block Model 113002/113002-2

5 lbs. (2.3 kg)

110020

6 tubes up to 25mm

4 Block Model 113004/113004-2

6 lbs. (2.7 kg)

110025

30 tubes up to 6mm

Power:

110035

Combination:

2 Block Model 113002

115VAC, 200 watts

3 tubes up to 25mm

2 Block Model 113002-2

230 VAC, 200 watts

5 tubes up to 13mm

4 Block Model 113004

115VAC, 380 watts

6 tubes up to 6mm

4 Block Model 113004-2

230 VAC, 380 watts

110040

1.5 microcentrifuge

Temp Range:

ambient +5°C to 120°C

tubes by 24 places

Uniformity:

±0.25°C @ 37°C

110045

0.5 ml tubes by 24 places

Stability:

±0.2°C from 37° to 100°C

110048

0.2 ml tubes by 48 places

Standard Block:

110051

96 well Microplate

Dimensions:

3” w x 3 ¾ ” d x 2” h

110096

0.2 ml Microtubes tubes

(7.6 x 9.5 x 5.1 cm)

110099

1.5 ml “dolphin” tubes 24 places

Shipping Weight:

2.0 lbs. (1.0 kg)

110002LID*

Lid for 110001 and

Net Weight:

1.0 lbs. (0.45 kg)

110002 Dry Baths

110004LID*

Lid for 110004 Dry Baths

*Temp Limited

Temperature Stability was determined by using NIST
traceable thermometers placed in thermometer hole in
standard 16MM aluminum blocks. Thermal grease was
used to ensure contact.


Temperature Uniformity was determined by using NIST
traceable thermometers placed in 16MM glass test tubes
with water in standard 16MM aluminum blocks. The
thermometers were suspended in the water. Six
positions were used in each block for the two-block
capacity unit and three positions were used in each
block for the four-block unit.


DEFINITIONS:

Uniformity is the variation in temperature, at the
location described above, in the block at any instant in
time.

Stability
is the variation in temperature at a single
point over a long period of time.

Note: The temperature in the test tube may be slightly
different than the temperature of the aluminum block.
The test tube material and the medium in the test tube
may affect the temperature.







Boekel Scientific

855 Pennsylvania Blvd.

Feasterville, PA 19053

Phone (215) 396-8200; (800) 336-6929

Fax: (215) 396-8264

e-mail: [email protected]

Advertising