Defibtech DDU-2450 Series Operating Guide User Manual
Page 5

DAC-E2530EN-BC
DAC-E2530EN-BC
8
9
For more detailed information, refer to the User Manual (on Defibtech User CD).
SETTING UP THE AED
The DDU-2000 Series AED is designed to be stored in a “ready”
state so that few steps are required to begin using the AED.
1
2
CONNECT
THE PADS TO
THE AED
INSTALL THE
BATTERY PACK
Ensure that the pads package has not expired.
Expired pads must not be used.
For more information refer to page 18 of this guide.
Ensure that the battery pack has not expired.
Expired battery packs must not be used.
When the battery pack is installed, the AED will turn
on and run a battery pack test. Wait for the test to
complete and for the unit to turn off.
For more information refer to page 19 of this guide.
3
4
CHECK THE
STATUS
READ THE
USER MANUAL
(CD)
For more information refer to page 16 of this guide.
With the AED off, press
and release the CENTER
softkey button. Ensure
that the AED Status
screen appears and that
the “AED Status” is “OK”.
AND
When the AED is off, the Active
Status Indicator (ASI) should flash
green. If the ASI flashes red, is
solid red, or if there is no flashing
light, the unit requires service.
Comprehensive information about the
DDU-2000 Series AED is found in the
User Manual (on Defibtech User CD).
AED Status screen
AED status
OK
Battery status
Expires 01/2018
Pads status
Adult
Expires 06/2018
AED S/N
099999999
Battery S/N
099999999
Software version
V2.2