CALIBRATE CELL

From the residual setup screen, press CELL to enter the cell calibration menu.

CALIBRATE CELL

ZRO SPAN

ESC

1

ZRO

Zero cell input

2

SPAN

Set the current CL

2

residual.

3

no function

4

ESC

Go back to the calibrate menu.

CALIBRATE ZERO SELECTION SCREEN

CALIBRATE ZERO

MAN AUTO

ESC

1

MAN

Manual zero

2

AUTO

Auto zero

3

no function

4

ESC

Go back to the calibrate cell menu.

MANUAL ZERO SCREEN

WAIT FOR...

0.0mV SET

ESC

The top line of the display scrolls the message, “ADD ZERO WATER & WAIT FOR STABLE mV.”
The left side of the bottom line shows the current cell mV.

1

No function

2

No function

3

SET

Zero cell input.

4

ESC

Go back to the calibrate zero menu.