Grant Instruments TBT-T100IN User Manual

Page 18

Advertising
background image

31082 V3

TBT-T100IN

Page 17

Operating Manual

www.grantinstruments.com

Display shows
“Open”

Faulty temperature
probe

Have a competent person check the probe for an
open circuit fault or contact Grant.

Display shows
“Shrt”

Faulty temperature
probe

Have a competent person check the probe for a
short circuit fault or contact Grant.

Unit showing
erratic
temperatures

Calibration values not
set correctly

Restore the factory calibration settings (see section
5.2.7) then re-calibrate if required (see section
5.2.6).

Display shows
“Hot”

Heating water at or
near to boiling without
lid

Heating very large
volumes of liquid with
large thermal losses

Add a lid to reduce thermal losses and leave unit
r

unning with the “Hot” warning present. If the alarm

has not cleared within 1 hour contact Grant.

Add measures to reduce thermal losses and leave
unit running with the “Hot” warning present. If the
alarm has not cleared within 1 hour contact Grant.

Display shows
“Fuse”

Faulty fuse(s)

Have a competent person check the internal fuses or
contact Grant.

Display shows
“Crct”

PCB fault

Have a competent person check the product or
contact Grant.

Display shows
“Self”

Safety relay fault

Have a competent person check the product or
contact Grant.

Stirrer motor not
rotating

Stirring propeller or
pump impeller is
obstructed

Faulty motor

Clear obstruction.



Have a competent person check the motor or contact
Grant.

Display shows
“Cold”

Accidentally entered
manufacturer reset
mode

Warning can be cleared by pressing S button, but
note that all settings (including any factory
calibration) will be reset to the default values.
Contact Grant Instruments service department for
assistance if required.


For any other errors or service requests, please contact Grant Instruments service
department.