KERN & SOHN FOB 5K1S User Manual
Page 9

9
FOB-S-BA-e-1311
Procedure when adjusting:
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
Operation
Start balance by pressing
The balance carries out an autotest and changes
then into the weighing mode, the zero display „0.0“
appears
(example)
(example)
Press
for approx. 3 seconds, „CAL“ appears
shortly, then „CAL 0“ and after that „LoAd“ is
displayed alternately with the value of the first
adjustment weight.
Place the first adjustment weight in the middle of the
weighing plate.
After a short time the value of the second
adjustment weight is displayed alternately with
„LoAd“
Remove the the first adjustment weight from the
weighing plate and place the second adjustment
weight in the middle of the weighing plate.
Adjustment occurs automatically after standstill
control.
„donE“ will be shortly displayed. Then the value of
the second adjustment weight will be displayed.
Adjustment process has been finished successfully.
The balance is now in weighing mode.