Installation instruction of scale
For IEX scales prepare an appropriate place for installation.
It should be dry, even and rigid
Acceptable operation temperature range -10°C to +40°C
The ground on which the platform is to be placed must be able to carry the
maximal load together with the platform
IEX scales are equipped with the indicator connected with the platform
The indicator can be mounted on the table or installed on a wall by means of a
handler.
Scale should be unpacked outside the hazardous area.
After unpacking remove all transport protections:
CAUTION!
In order to dissipate static electricity, it is necessary to ground the
scale’s indicator.
Places for plugging the wires are marked with “ ” symbol.
Before moving the scale to its intended place of use, prepare a conductor for
electrostatic grounding of the indicator.
The grounding conductor should be ended with a loop that is screwed to the
indicator using M4 nut.
In order to ensure constant holding down the loop to the indicator´s housing, use
tooth lock washer.
The grounding conductor should have yellow and green colour with minimal
intersection 4mm².