KERN & SOHN KFA 6000V20L User Manual

Page 13

Advertising
background image

KFA V20-IA-e-1312

13

8.5 Instant help
In case of an error in the program process, briefly turn off the weighing system and

disconnect from power supply. The weighing process must then be restarted from

the beginning.

Help:
Fault

Possible cause

The displayed weight is permanently

changing

Draught/air movement
Floor vibrations
Weighing beams come into contact with

foreign substances.

Electromagnetic fields / static charging

(choose different location/switch off

interfering device if possible)

The weighing result is obviously

incorrect

No zero display for unloaded weighing

beams.

Adjustment is no longer correct.
Great fluctuations in temperature.
The weighing beams are not level.
Electromagnetic fields / static charging

(choose different location/switch off

interfering device if possible)

Should other error messages occur, switch balance off and then on again. If the error

message remains inform manufacturer.