KERN & SOHN VHE 2T5A User Manual
Page 13

Press
for approx. 3 sec., the count down for setting
the second adjustment point runs off automatically.
The weight for the second adjustment point is displayed.
Moreover two points appear in the display.
Place 2000 kg
After placing the weight the weighing system leaves the
reference height, the display flashes
Place the weighing system on reference height, the display
stops flashing.
Press
for approx. 3 sec., the count down for setting
the third adjustment point runs off automatically, 0 is
displayed
Remove weight.
Press
, vU004 is displayed, followed by PAr
Press
for approx. 3 sec., SAVE is displayed,
followed by 0.
The balance is now in weighing mode.
Adjustment now has finished.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.