8.2
Carry out linearization
Turn off balance and attach suitable lifting tackle.
Turn on balance with the lifting tackle attached and press
during self-test. The first function „F0bk“ is displayed.
Press
repeatedly until „F5CAL“ is displayed.
Press
, the last parameter set will be displayed.
• „nonLi“
Adjustment of balance
or
• „LinE“
Linearisation of balance
Select „LinE“ by
Press
, “LoAd 0“ will appear. Make sure that no loads
apart from the lifting tackle are attached to the hook.
Wait for stability display and press
again. “LoAd 1“
(1/3 max) will be displayed.
Attach first adjustment weight, wait for stability display
Press
, “LoAd 2“ (2/3 max) will be displayed.