11.4 Error messages
Error
message
Description
Err 0
Value outside the A/D changer range
Err 2
Weighing value
≤ Zero range
Err 6
Internal solution
≤ 0.15 µV/d range
Err .
Wrong password
E 1
Internal value < zero range
E 2
Internal value > zero range
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.