10 Instant help


In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.

Help:

Fault Possible

cause

The displayed weight does
not glow.

• The balance is not switched on.

• Batteries are inserted incorrectly or empty

• No batteries inserted.

The displayed weight is
permanently changing

• Draught/air movement

• Table/floor vibrations

• The weighing plate is in contact with foreign bodies

or is not correctly positioned.

• Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The weighing result is
obviously incorrect

• The display of the balance is not at zero

• Adjustment is no longer correct.

• Great fluctuations in temperature.

• Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.