Switch the MAINS on then Turn the simmerstat/thermostat or temperature controller up using the up or

 

 

 

 

 

 

 

   

 

 

 

 

 

   

down keys only as required. 

4.

An indicating lamp ‘OP1’ will flash when the heaters are operating. 

5.

Ensure the manual reset over heat cut off (brown button below the controller) is set by pressing it in or that

 

 

 

 

 

 

   

 

 

 

 

   

   

       

 

the over temperature alarm AL4 is set above the desired operating temperature (depending on which is

 

 

 

 

   

 

 

 

 

 

 

 

 

   

present). 

 
 
Safety Overtemperature cutoff ­ manual reset button 
If the water bath is fitted with a manual reset button, be sure this is depressed before use. If the water level is too

 

 

 

   

 

   

 

 

 

 

 

   

 

 

   

 

 

   

 

low or runs out the switch will pop out and turn the heaters off until the reset is depressed. 
 
If the water bath is fitted with an inbuilt over heat safety alarm it must be set to slightly above the desired setpoint 
temperature and will prevent overheating. Change the ALM4 value to slightly above the setpoint temperature as set 
out in the instruction page. The Alarm will be activated if the bath reaches the ALM4 value or if the sensor is 
broken or damaged.  To restart the bath to heat, acknowledge the alarm by pressing the page and scroll keys 
simultaneously (left two keys on key pad together for 1 second only).  
 
 
Maintenance 
The casing is finished with stainless steel and to maintain appearance should be wiped over with a non abrasive

 

   

 

 

 

 

   

 

 

 

 

 

 

   

 

 

cleanser. 
The interior is also stainless steel and may also be cleaned with a solvent. 
 
 
Trouble Shooting 
 

SYMPTOM 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check the water bath is plugged in and

 

 

 

   

   

 

power switched on. 
 

2. Ensure mains power supply point is

 

 

 

 

 

 

functioning by using a test appliance on

 

 

   

 

 

 

power socket. 
 

3. Check internal RCD has not tripped. 

Failure to heat or maintain temperature 
 
(Indicator light is on) 

1. Ensure the temperature controller set

 

 

 

 

 

point is above ambient. 
 

2. Check the safety controller fitted is above

 

 

 

 

   

 

the main controller setting. 
 

3. Cut out button has popped out (if fitted). 

 
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the unit from the

 

 

 

 

 

 

 

 

   

 

 

 

 

 

   

 

 

 

 

manufacturers label. 
 
 
Declaration of Conformity 
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications provided, but to

 

   

 

 

 

   

 

 

 

   

 

 

 

 

   

also meet Australian Electrical Standard AS3820 and EMC Standard AS/NZ1044:1995,and therefore being

 

 

 

 

 

 

 

 

 

 

 

 

accredited with a C Tick label. 
 