Operation 
1.

Connect the bath to an alternating current supply of voltage specified on the rating plate. 

2.

Ensure the oil level is above the elements.  

 
DANGER: THIS BATH MUST NOT BE CONNECTED TO A DIRECT POWER SUPPLY 
 
3.

Switch the MAINS on then Turn the simmerstat/thermostat or temperature controller up as

 

 

 

 

 

 

 

   

 

     

required and check the indicating lamp is illuminated. 

4.

An indicating lamp on the controller will illuminate when the heaters are operating. 

5.

Ensure the over heat cut off is set by pressing it in or the overtemperature alarm AL4 is set

 

 

 

 

 

   

   

       

 

 

 

   

 

above the desired operating temperature (depending on which is present). 

 
 
 
Safety Cut/ Overtemperature 
If the bath is fitted with a manual reset button, be sure this is depressed before use. If the level is

 

 

   

 

   

 

 

 

 

 

   

 

 

   

 

   

too low or runs out the switch will pop out and turn the heaters off until the reset is depressed. 
 
If the bath is fitted with an inbuilt over heat safety protection. It must be set to slightly above the 
desired setpoint temperature and will prevent overheating. It will maintain the Alarm Set Value. 
Change the AL4 value to slightly above the setpoint temperature as set out in the instruction page. 
The Alarm will be activated if the incubator exceeds the AL4 value or if the sensor is broken or 
damaged. Adjusting the Alarm Hysterisis value will cause the incubator to cool down by that value 
before it begins heating again. For example if the Alarm Hysterisis is set to 100 the incubator must 
cool down by 100°c below AL4 before it will begin to reheat. Where you require the incubator to 
switch off completely should an alarm condition result then please contact the manufacturer to 
purchase the OST100 manual overtemperature reset control system. 
 
 
Maintenance 
The casing is finished with stainless steel and to maintain appearance should be wiped over with a

 

   

 

 

 

 

   

 

 

   

 

 

   

non abrasive cleanser. 
The interior is also stainless steel and may also be cleaned with a solvent 
 
 
 
 
 
 
 
 
 
 
 
 
 
 