LABEC M-SCEN-206 User Manual

Page 8

Advertising
background image

4. Operation Panel:

(01) START: After setting time and speed, press it and rotor rotates immediately. .

(02) STOP: For emergency stop.

(03) BRAKE: It is with auto-braking function, when presetting time is due.

(04) OPEN: It is used at speed under 100rpm.

(05) RPM: Press【RPM】(flash), adjust ▲ or ▼ then press “RPM” to finish it.

(06) TIME: Press【TIME】(flash), adjust ▲ or ▼ then press “TIME” to finish it.

(07) ▲ INCREASE: to add speed or time.

(08) ▼ DECREASE: to reduce speed or time.

(09) RECALL: to display settled speed and time.

(10) STORE: to remember settled speed and time.

After setting speed and time, press【STORE】twice. If you need presetting function,

press 【STORE】first then press【PROG】to select demand figure then press【STORE】.

(11) PROG: to display speed and time.

Press【RECALL】(flash) then press “PROG” to select demand figure, then press

【RECALL】again to confirm.

(12) DOOR LAMP: when cover is open, it lights up.

(13) BRAKE LAMP: when it lights up, it will decrease speed automatically.

(14) IMBAL: when it lights up, it is imbalanced on rotor and stop running. You have to

press【OPEN】to open cover and check problem.

(15) RUN: it lights up during rotation.

(16) DISPLAY: to show speed, time and settled figure.

(17) Transient centrifugal function: it can reach specified speed in transient time without

speed and time limited.

6