Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01

12. Trouble-Shooting

Problem Description

Possible Causes

Solutions

Non-running of unit, no
display of LED indicator

Failure of power supply

Proper wiring of the unit

Damage of fuse

Replace the fuse

No display of LED indicator,
however the unit is running

High frequency interference
from other equipments

Use dedicated line for the
unit

Wide fluctuation of
temperatures

Refrigeration parameters not
correctly set

Reset the parameters

Measured temperature is
above preset value
Frequent overtemperature
alarming

Preset temperature value too
high

Remove the wind inlet
cover at the back of the
unit or set the
refrigeration value to 0.5


High Noise Level

Unit is not placed on flat floor Place the unit on flat floor
Platform unsecure

Fasten the four knobs on
the platform

Flask clamps unsecure

Fasten the nuts of flask
clamps





13. Packing List

Upon receipt of the unit, examine if there is any damage done to the unit during
transportation. In the event of shipping damage, a claim must be filed with the
transportation carrier within 7 days of receipt.

Remove the packing material from the box, take care that the accessories are not
thrown away with the packing material. Lift the unit out of the box. Do not lift the
unit by the door handle!
Check all the items have been received. Included in the
box should be the following:

Item

Quantity

Main unit

1

Platform

2

Power cord

1

Fuse

2

Tool for changing platforms

1

Water tray

1

Instruction Manual

1