Temperature Controls 
See Appendix A 
 
Maintenance 
The cabinet is finished with painted epoxy and to maintain appearance, should be wiped over 
with a non abrasive household cleanser.  The chamber is manufactured from formed plastic and 
may be cleaned with a gentle cleaner.  The door is also plastic lined and should only be cleaned 
with general purpose cleaners. 
If a fan motor is fitted the fan motor is pre­lubricated with sealed bearings which should not 
require maintenance for some time depending upon extent of usage.  The shelves are 
manufactured from plastic coated steel and a non abrasive cleaner should be used.  The 
magnetic door gasket should be cleaned with detergent only, ensuring it is dried completely 
after washing. 
 
Defrosting 
Over a period of time the refrigerator’s cooling coil will build up a layer of frost which will reduce 
the operation of the unit.  We recommend defrosting when the ice frost layer is around 5mm 
thick or when the refrigerator starts to slowly heat up in temperature. Turn the unit off for a 
period of 24 hours with the door opened to allow the frost to melt away.  This should be done 
every 3 months or as required.

 

 
Trouble Shooting 
 

SYMPTON 

REMEDY 

No Power 
 
(Indicator Light is off) 
 
 
 
 
 
 

1. Check pharmacy vaccine refrigerator is 

plugged in and power switched on. 
 

2. Ensure mains power supply point is 

functioning by using a test appliance on 
power socket. 
 

3. Check internal RCD has not tripped. 

 

 

Failure to maintain temperature 
 
(Indicator light is on) 

1. Ensure the temperature controller set 

point is above ambient. 
 

2. Check the safety controller fitted is above 

the main controller setting. 

 
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of 
the unit from the manufacturer’s label. 
 
 
Declaration of Conformity 
Each product is thoroughly inspected and tested to not only ensure that it meets the 
specifications provided, but to also meet Australian Electrical Standard AS3820 and EMC 
Standard AS/NZ1044:1995,and therefore being accredited with a C Tick label. 
 

 