4. Temperature Calibration________________________

 

 

The Labec Warming Tray has been calibrated as a set.  Warming 

Tray can be calibrated by placing a calibrated thermometer on 

the plate for 1 hour and then measuring the difference. Altering 

the “PV0fS” value to match is then carried out. (see the 

Appendix A manual for Calibrating and only perform this if more 

than 2ºc difference is recorded)

 

 
 

5. Troubleshooting Guide__________________________

 

 

Many operating problems may be solved by reading and carefully 

following the instructions in this manual.  Some suggestions for 

troubleshooting are given below.  Should these suggestions not 

resolve the problem, contact the SERVICE DEPARTMENT on 

02-9560 2811 for assistance or email to [email protected].  Please 

supply the serial number and include a full description of the 

problem.

 

 
 

Problem                                                              Comments______ 

 

LED do not light up                             1. Check the RCD

 

                                                                

2. Ensure that the AC 

 

                                                                    

power switch is ON.

 

                                                                

3. Check the three-pronged

 

                                                                     

power cord is properly

 

                                                                     

plugged into a grounded 

 

                                                                     

three-prong AC outlet 

 

                                                                     

of the appropriate 

 

                                                                     

voltage. 

 
 

 

Laboratory Equipment Pty Ltd - Warming Tray 2014 Ver2.01 Instruction Manual 