Specifications product specifications, Environmental conditions, Electrical specifications – Oakworks 300 Series Procedure Chair Rev.D User Manual
Page 17

14
SPECIFICATIONS
PRODUCT SPECIFICATIONS
Weight
280-325 lbs. (127-148 kg.)
Shipping Weight
372-417 lbs. (169-189 kg.)
Lifting Capacity
550 lbs. (250 kg.)
ENVIRONMENTAL CONDITIONS
Conditions
Temperature
Humidity
Atmospheric Pressure
Normal Use
50° (10°C) to 104° (40°C)
20% to 60% RH
98 to 105 kPa
Storage & Transport
-20° (-29°C) to 135° (57°C)
20% to 95% RH
98 to 105 kPa
ELECTRICAL SPECIFICATIONS
Designed for:
North America
Europe
Input Service
120 VAC/15 amp/60 Hz
230 VAC/10 amp/50/60 Hz
Current Draw
5.8 amps
2.9 amps
Maximum Momentary
Current Consumption
6.0 amps
3.0 amps
Voltage Output to Actua-
tors
24 VDC
24 VDC
Electric Shock
Protection
Class 1 Equipment
Class 1 Equipment
Tabletop Applied Part
Type B Applied Part
Type B Applied Part
Ingress Protection
Rating
IPX0
IPX0
Made of Operation
Intermittent Operation MAX 2
minutes ON
18 minutes off
Intermittent Operation MAX
2 minutes ON
18 minutes off
Fuse Specifications
M 10AL 125V
M 10AL 125V