Unit will not heat up at all

1/ verify that controller is asking for heat by looking for controller light

– if

pilot light is not on continuously, there is a problem with the controller.
2/ check amperage

– amperage should be virtually at maximum rated

(data plate) amperage.
3/ do all controller functions work?
4/ is the Thermostat set high enough?

– for diagnostics, should be fully

clockwise with the pilot light never on.
5/ has the fuse/circuit breaker blown?

Indicated chamber temperature unstable

1/ ±0.1 may be normal
2/ is fan working?

– remove top panel and verify movement of cooling fan

in center of shadow box
3/ is ambient room temperature radically changing

– either door opening

or room airflow from heaters or air conditioning ?

– stabilize ambient

conditions.
4/ sensor miss-located, damaged or wires may be damaged - check
mounts for control and Thermostat sensors, then trace wires or tubing
between sensors and controls.
5/ calibration sensitivity

– call Customer Service

6/ Overtemperature set too low

– be sure that the thermostat is more

than 5 degrees over desired set point; check if OVERTEMP pilot is on
continuously; turn controller knob completely clockwise to see if problem
solved then follow instructions in owner’s manual for correct setting – see
section 5.7
7/ electrical noise

– remove nearby sources of RFI including motors,

arcing relays or radio transmitters.
8/ bad connection on temperature sensor or faulty sensor

– check

connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage.
9/ bad connections or faulty relay

– check connectors for mechanical

soundness and look for corrosion around terminals or signs of arcing or
other visible deterioration.

Will not maintain set point

1/ assure that set point is at least 5 degrees over ambient -
2/

SEE IF AMBIENT IS FLUCTUATING

Display and reference thermometer don’t
match

1/ calibration error

– see section 5.6

2/ temperature sensor failure

– evaluate if pilot light is operating

correctly.
3/ controller failure

– evaluate if pilot light is operating correctly

4/ allow at least two hours to stabilize.
5/ see if reference thermometer is certified.

Can't adjust set points or calibration

1/ turn entire unit off and on to reset.
2/ if repeatedly happens, call Customer Service

Calibrated at one temperature, but not at
another

This can be a normal condition when operating temperature varies widely.
For maximum accuracy, calibration should be done as close to the set
point temperature.

REFRIGERATION

Temperature can't get up to set point

1/ assure that power is going to heating coils.
2/ if the displacement is erratic, see if air is being circulated.
3/ confirm that controller is calling for heat (check front panel light).
4/ if light not coming on, check control set point and Overtemperature set
point .
5/ confirm that fan is operating and airflow is not blocked.
6/ reset by turning unit off and on.