P

OWER

S

OURCE

This oven is intended for a 230 volt, 20 Amp, 50/60 Hz application. Check the

oven’s data plate to

verify voltage and ampere requirements before connecting to a power source.

The

oven’s power feed must be wired into an earth grounded source.

The power source for the oven must conform to all national and local electrical codes Supplied
voltage must not vary more than 10% from the data plate rating. Damage to the oven may
result if supplied voltage varies more than 10%.

Before energizing the oven it must be earth grounded using the protective conductor terminal
(green with yellow stripe wire. Do not remove the protective conductor (earth connection).
Removing the protective conductor will negate the oven’s protections against potentially dangerous
electric shocks and create a possible fire hazard.

A separate circuit is recommended to prevent possible loss of product due to overloading
or failure of other equipment on the same circuit.

A switch or circuit-breaker must be used in the building installation.

The required circuit-breaker is 20A

The wires for power source connection should be in accordance with the following:
SVAC9-2 (VPX9-2): Green/Yellow

– Earth; Red – Hot; Black – Hot.

The Disconnect must be in close proximity to the unit and within easy reach of the operator.

The Disconnect must be marked as the disconnecting device for the equipment.

L

EVELING

Make sure that the oven is level and stable. Each SVAC9-2 oven ships with four leveling feet. Insert
one leveling foot into each of the four holes in the bottom corners of the oven. Stand the oven
upright. Then, adjust the foot at each corner until the oven stands level and solid without rocking.
To raise a foot, turn it in a counterclockwise direction; to lower a foot, turn it in a clockwise direction.

Note: To prevent damage when moving the oven, turn each of the four leveling feet completely

clockwise.