Appendix d: recommended – Soniclean S2800 User Manual
Page 25

Appendix D: Recommended
S-2800/S-2800MRS disinfection protocol
It is Soniclean’s protocol to disinfect every irrigator supplied for trial at the end of every trial and
immediately prior to commencing the next trial. Soniclean or their distributors/agents will ensure this is
undertaken unless the trial organisation prefers to conduct the disinfection themselves.
The protocol to be followed:
1. Drain irrigator and turn off at power point.
2. Apply protective gloves (disposable non sterile latex gloves are suitable).
3. Remove tray; thoroughly wipe down all visible surfaces of the irrigator including the lid, tank, all
outside surfaces, hoses and attachments with disinfectant solution
1
.
4. Remove detergent hose and replace with disinfectant hose connected to bottle with slightly more
than required amount of disinfectant
1.2
. Insert tray.
5. Turn irrigator on at power point whilst holding down the program key (this will allow you to re-
program the parameters).
6. Change soap % parameter to required amount of disinfectant. Ensure irrigate function on.
7. Press “clean”. The irrigator will now add appropriate amount of disinfectant to tank and in doing so
will disinfect the internal plumbing fittings as well as irrigator ports.
8. Once clean is complete, disconnect disinfectant bottle and replace with bottle filled with clean
water.
9. Press “shift” and “refill” keys, hold for three seconds. This will cause irrigator to empty of water and
disinfectant and then to fill with water. This will rinse disinfectant out of the unit.
10. Repeat steps 5 and reprogram soap parameter to factory setting (i.e. 1.2)
11. Dry all external surfaces and pack unit ready for transport.
___________________________________________________________________
Note regarding Disinfectant solutions: The recommendation is to use a TGA approved medical grade
disinfectant which is recommended for use in hospital settings for disinfecting horizontal and vertical
working surfaces.
Soniclean currently uses a product called LCC (supplied by Dominant Chemicals). For disinfecting the
S-2800 Irrigator Operating Manual
Page 21
© 2013-2014 Soniclean Pty Ltd
Version 20
March 2014
KEY TO PROGRAMMING DISPLAY
PARAMETER
DISPLAY
MACHINE
MIN
MAX
FACTORY
SETTINGS
Clean time
Ct
mins
1
99
10
Ultrasonic power level
PO
1
3
3
No. of cleans between
drains
Cn
0
99
10
Detergent concentration
SP
percentage
0.0
5.0
0.5
Start up degassing time
G1
mins
0=Off
99
10
Degassing time prior to
each clean
G2
mins
0=Off
99
2
Test time
Tt
sec
0=Off
99
30
Irrigate default
Id
0=Off
1=On
1
Irrigate rest periods
Ir
1 rest period
= 5 seconds
5 period
(5 x 5 = 25
seconds)
50
5