ThermoWorks PHARMALARM 2 MAX/MIN RECORDING THERMOMETER W/CAL CERTIFICATE User Manual

Pharmalarm™ 2, Nist-traceable max/min recording thermometer

Advertising
background image

PharmAlarm™ 2

NIST-Traceable MAX/MIN Recording Thermometer

Instrument Operation

The PharmAlarm 2 is a NIST-Traceable two-channel alarm thermometer that records

the date and time of Max and Min readings for both channels. The left display shows

current temperature information on the selected channel. The right display shows time

and date information.
When the reading is higher or lower than the programmed alarm limit, the alarm on the

PharmAlarm 2 will sound, HI or LOW will flash, and the red light will blink. Pressing

any button will stop the audible alarm. The HI or LO indicator will flash until the reading

is within limits. The red light will continue to blink to indicate that an alarm has been

triggered.

Setup and Installation

1. Open the battery cover and install the batteries observing the correct polarity.

Replace the battery cover.

2. Select the desired temperature unit (°F or °C) using the slide switch on the back.

3. Fully extend the probe wires.

4. Attach the thermometer using the magnets on the back or the flip out desk stand.

5. Position the sensors in the desired location.

6. Peel the clear protective sheet from the LCD display.

Time and Date Display

1. Press TIME/DATE to switch the display to show the current time or date.

2. The date display will switch back to time display if idle for 10 seconds.

Set the Time

1. In time display mode, press and hold SET for about 3 seconds to enter time setting

and the hour digits will flash.

2. Press

UP

to adjust the digits to current hour.

3. Press SET once to confirm and start minute setting; the MIN digits will will flash.

4. Press

UP

to adjust the digits to current minute.

5. Press SET once to confirm and start second setting; the SEC digits will will flash.

6. Press

UP

to adjust the digits to current second.

7. Press SET again to finish time setting and return to normal display.

Set the Date

1. In date display mode, press and hold SET for about 3 seconds to enter date setting

and the year digits will will flash.

2. Press

UP

to adjust the digits to the last two digits of current year.

3. Press SET once to confirm year setting and then start month setting; the Month

digits will will flash.

4. Press

UP

to adjust the digits to current month.

5. Press SET once to confirm month setting and then start date setting; the Date digits

will will flash.

6. Press

UP

to adjust the digits to current date.

7. Press SET again to finish date setting and return to normal display.

Switch Channel Display

Press S1/S2 once to switch the display for channel S1 or S2 reading.

Maximum/Minimum Reading Memory

1. Press MODE once to display the minimum measured reading of the selected channel

and the time display will show when this minimum reading was recorded.

2. Press TIME/DATE to show the date for the minimum measured reading.

3. Press MODE once to display the maximum measured reading and the time display

will show when this maximum reading was recorded.

4. Press TIME/DATE to show the date for the maximum measured reading.

S. Press MODE three times to return to normal display.

6. To clear the minimum and maximum reading memories, press CLEAR once during

displaying the minimum or maximum reading. All readings of both channels and

time stamps will be reset to current readings and time.

7. Clear the memory once before taking new minimum or maximum readings.

Low/High Alarm Setting

1. Select the desired channel, S1 or S2.

2. Press MODE three times to enter low/high alarm setting and the LO icon will be on.

3. Press and hold SET for about 3 seconds and the LO digits will will flash.

4. Press

UP

to input the desired low limit.

5. Press SET to confirm low alarm limit and start high alarm limit setting. The HI icon

will be on and the digits will will flash.

6. Press

UP

to input the desired high limit.

7. Press SET to confirm high alarm limit and finish the setting.

8. The alarm will sound when the reading is lower or higher than the alarm limit. Press

any button to stop the alarm sound. The HI or LO icon will continue to flash until the

reading is within the limits.

Alarm On/Off

Press

%

to switch alarm limit off or on.

Alarm Indicator

When the reading exceeds the alarm limit the red alarm LED will continue flashing to

indicate that an alarm has been triggered until the alarm is cleared. Press and hold

CLEAR for about 3 seconds to cancel.

Note

1. Do not operate the thermometer in the environmental temperature lower than 32°F

(O°C) or higher than 122°F (5O°C) otherwise incorrect readings or damage to the

thermometer may result.

2. If the thermometer is not in use for a long period of time then remove the batteries

from battery compartment to avoid battery leakage.

Specifications

Measuring Range -58 to 158°F (-50 to 70°C)

Display Accuracy ±1.8°F (1°C)

Display Resolution 0.1°

Display Reading Update 10 seconds

Program Alarm Resolution

Time Accuracy ±3 seconds per day

Time and Date Display Format 12-hour and month/day/year format

Effective Calendar Period 2000 to 2099

Battery 1.5 volt, type AAA or equivalent x 2 pieces (included)

Probe Cord Length 39.4 inches (1,000 mm)

Display Size 0.63 H x 3.5 W inches

(16 H x 90 W mm)

Product Size 1.77 H x 5.3 W x 0.9 D inches

(45 H x 135 W x 23 D mm)

Included Accessories 1) 1.5 volt, type AAA battery x 2 pieces

2) 2 sensor vials filled with glass beads

3) NIST-Traceable Calibration Certificate

Utah, U.S.A.

Ph: 801-756-7705

Fax: 801-756-8948

www.thermoworks.com

Made in China

PharmAlarm™ 2

MAX/MIN Recording Thermometer

P-11-035-01-ab