Lift diaphragm carefully sidewise.

+

Never use a pointed or sharp-edged tool to lift the dia-

phragm.

➨

Use the diaphragm key to grip the diaphragm support disc

below the diaphragm.

➨

Press down the diaphragm clamping disc to bring the

diaphragm to its lowest oscillating position. Unscrew dia-

phragm support disc with diaphragm and diaphragm clam-

ping disc.

+

If the diaphragm is difficult to separate from the diaphragm

support disc, immerse assembly in naphtha or petroleum

ether. Do not inhale vapors!

+

Check for washers between the diaphragm support disc

and the connecting rod. Make sure that the original num-

ber is reassembled.

➨

Note position of valves.

➨

Unscrew four screws at the pump head, paying attention

to washers.

Remove housing cover.

➨

Unscrew union nut and remove hose from elbow fitting.

➨

Remove housing cover insert with elbow fitting.

➨

Remove head cover and valves.

+

Check diaphragm for damage.

+

Attention: In case of a defect diaphragm the interior

of the pump housing has to be decontaminated, e. g.,

using a spray disinfection.

+

The appropriate decontamination method has to be

ensured by the user.

+

Never remove parts by using a pointed or sharp-edged

tool (e.g., screw driver), we recommend to use a rubber

mallet or compressed air (to be blown carefully into port).