Shutdown

Short-term:

Empty bottle.

Has the pump been exposed to

condensate?

Set the VHC

pro

to continuous aspiration and the BVC to

maximum underpressure and allow the pump to continue

to run for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits? Check and clean

pump head if necessary.

Long-term:

Take measures as described in short-term shutdown sec-

tion.

Close inlet and outlet port (e. g., with transport caps).

Store the pump in dry conditions.

The pump achieves its

pumping speed, ultimate vacu-

um and vapor pumping rate only at operating temperature

(after approx. 15 minutes).

Prevent internal condensation, transfer of liquids or dust.

The diaphragm and valves will be damaged, if liquids are

transferred through the pump in significant amounts.

In case of excess temperature the motor is shut down by a

thermal cutout in the winding. Manual reset is necessary.

Switch off or unplug the pump. Identify cause of failure

and eliminate. Wait approx. five minutes before restarting

the pump.

Attention: In case of supply voltage below 100V (100-

120V version), the lock of the cutout might be restricted

and the pump might restart on its own after sufficient cool-

ing down. Take suitable precautions if an automatic restart

of the pump may lead to a critical dangerous situation.

After a power failure, the system will automatically start

again.