Cleaning and decontaminating, Notice – VACUUBRAND 8-channel tip adapter with ejector User Manual
Page 18
Advertising

Prior to autoclaving set the VHC
pro
to
“autoclaving”.
Prior to autoclaving loose the cap.
Cleaning and decontaminating
Autoclaving
The VacuuHandControl VHC
pro
with adapters
and aspiration hose is designated for steam
sterilization at 121°C and 2 bar absolute (1 bar
overpressure). Time of exposure according to
DIN 58946 t
e
= 20 minutes.
Ensure not to exceed a temperature of 121°C.
Comply with instructions for use of the manu-
facturer of the autoclave.
It is the user´s responsibility to ensure effective
autoclaving.
NOTICE
NOTICE
After autoclaving tighten the nose cones if nec-
essary with the mounting tool (enclosed in de-
livery), see section „Replacing the nose cones“.
NOTICE