If the automatic mode has been preselected “End” is displayed.
☞ Switching between automatic switching off activated or not
activated is possible by pressing key
▲ or ▼.
➨ Confirm by pressing key START/STOP.
☞ The controller switches to normal operation mode (process
control not active).
☞ Automatic switching off as soon as the controller detected
that evaporation is completely finished. The pump is switched
off, the cooling water is running on, if a cooling water valve
is preselected.
☞ Without automatic switching off, the pressure in the vacuum
system is adjusted after evaporation is completed if neces-
sary until the ultimate vacuum of the system.