Replacing diaphragms and valves, See section „re, Placing diaphragms and valves“, pg. 99) – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual
Page 99
page 99 of 120
Replacing diaphragms and valves
+
Please read section ”Replacing diaphragms and valves” com-
pletely before starting maintenance.
The pictures may show other versions of pumps. This does not change
the method of replacing diaphragms and valves.
➨
Never operate the pump if covers or other parts of
the pump are disassembled.
➨
Before starting maintenance,
disconnect the electri-
cal power cord. Wait two minutes after isolating the
equipment from AC power to allow the capacitors to
discharge.
➨
Ensure that the pump cannot be operated accidentally.
➨
Note: The pump might be contaminated with the pro-
cess chemicals that have been pumped during opera-
tion. Ensure that the pump is decontaminated before
maintenance.
➨
Avoid the release of pollutants.
+
Never operate a defective or damaged pump.
+
Take adequate precautions to protect people from the
effects of dangerous substances that may have con-
taminated the pump and may be released upon dis-
assembly. Ensure that the maintenance technician is
familiar with the safety procedures which relate to the
products processed by the pumping system.
Use appropriate protective clothing, safety goggles
and protective gloves.
+
Allow sufficient cooling of the pump before starting
maintenance.
+
Vent the pump and isolate it from the vacuum sys-
tem before you start maintenance.