fuse holder
➨
Keep the snap-fits squeezed and pull the fuse holder out.
➨
The fuse holder contains two fuses of the same type. Replace the
defective fuse by a fuse of the same type (see “Technical data”, pg.
➨
Insert the fuse holder into the housing until it snaps.
Replacing the device fuse
➨
Switch off the pump.
 
➨
Disconnect the electrical power cord before remov-
ing the fuse holder.
Identify and eliminate the cause of failure before switch-
ing on the pump again.