Calculation example for the connection of an intrinsically safe circuit to the DVR 4S

Do not exceed the max. permitted capacity and inductance of the intrinsically safe circuit (according to
corresponding standards).

1 Maximum values for the intrinsically safe circuit defined in the standard (depending on connected

components):
C

max

= 110 nF

L

max

= 5 mH

2 Subtract the effective maximum capacities and inductances of the connected devices, e. g. DVR 4S:

C

max´

= 110 nF - 64 nF = 46 nF

L

max´

= 5 mH - 60 µH = 4.9 mH

3 Calculation of the permitted cable length:

Assumed values of the cable parameters in case of non-shielded cable:
C

= 110 nF/km

L = 0.65 mH/km

Calculation with regard to the capacity:
l = (46 nF/110 nF) x 1000 m = 418 m

Calculation with regard to the inductance:
l = (4.9 mH/0.65 mH) x 1000 m = 7500 m

4 The smaller value limits the cable length:

l

max

= 400 m