Safety information, General information, Intended use – VACUUBRAND VSP 3000 User Manual

Page 7: Connecting the pressure transducer, Notice

Advertising
background image

page 7 of 18

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999159 / 22/07/2014

General information

Read and comply with this manual before installing or operating the equip-

ment.

Do not use the equipment if it is damaged.

To operate the pressure transducer VSK 3000 a VACUUBRAND vacuum gauge

DCP 3000 or a vacuum controller CVC 3000 or VNC 2 is required. To operate the

pressure transducer VSP 3000 a VACUUBRAND vacuum gauge DCP 3000 or a

vacuum controller CVC 3000 is required. Remove all packing material, remove the

product from its packing-box, remove the protective covers and keep, inspect the

equipment.

If the equipment is damaged, notify the supplier and the carrier in writing within three

days; state the item number of the product together with the order number and the

supplier’s invoice number. Retain all packing material for inspection.

Intended use

Ensure that the individual components are only connected, combined and operat-

ed according to their design and as indicated in the instructions for use. Use only

original manufacturer’s spare parts and accessories. Otherwise the safety

and performance of the equipment, as well as the electromagnetic compatibility

of the equipment might be reduced.

The pressure transducers are designed for

ambient and gas temperatures at

the measurement connection of +10°C to +40°C at continuous operation or of up

to +80°C for short periods (<5 minutes). Ensure that the maximum permitted gas

temperature at the pressure transducer (see ”Technical data”) is not exceeded.

Use the equipment and all system parts for the intended use only, i.e. for measure-

ment and control of vacuum in vessels designed for that purpose.

Connecting the pressure transducer

Avoid uncontrolled overpressure (e. g. when connecting to a locked or blocked

tube system).

Risk of bursting.

Comply with max. permitted pressure at the pressure transducer, see section

”Technical data”.

• Connect hoses at the pressure transducer gas tight.

• Ensure stability of the hose connections.

• Check that supply voltage and current conform with the equipment (see rating

plate).

• When working with residues, aggressive or condensable media, install a gas

washing bottle if necessary.

NOTICE

NOTICE

Safety information

Advertising
This manual is related to the following products: