Data output mode – A&D Digital Platform Scales FG-60KAM User Manual

Page 20

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

Data output mode

□ stream mode Function setting “F5-0"

The scale outputs the current display data. The data-update rate is approximately 10
times per second. This rate is the same as the display-update.

The scale does not output data while it is in setting mode.

□ Print switch mode

When the

Function setting “F5-2”

PRINT switch is pressed while the weighing data is stable (STABLE

indicator is on), the scale transmits the data.

□ Auto-print mode +/- data Function setting “F5-3“

The scale transmits the weighing data when the display is stable (STABLE indicator

is on) and the data is more than +4d or iess than -4d of weight data,

d = minimum weight display (see “12-1 Specifications”)

When in counting mode, “d" is equal to minimum weight display of kg mode.

The next output can be obtained after the display returns to between -4d and +4d.

□ Auto-print mode + data Function setting “F5-4“

The scale transmits the weighing data when the display is stable (STABLE indicator

is on) and the data is more than +4d of weight data.

d = minimum weight display (see “12-1 Specifications”)

When in counting mode, “d” is equal to minimum weight display of kg mode.

The next output can be obtained after the display returns to below +4d.

□ Command mode Function setting “F5-1

In the command mode, the scale is controlled by commands that come from the

personal computer and so on.

Command list

“Q” command Command to request the current weighing data.

Command

Reply

Q

C

r

L

f

S T

1

+ 001

“Z” command Same operation as the

Command

Z

C

r

L

f

□ This command works as

RE-ZERO

4 5 _ k g C

r

L

f

or

ZERO switch.

and as ZERO for U.S.A. modeis.

“T” command Same operation as the TARE switch

Command T

C

r

L

f

□ This command works as TARE for U.S.A. models.
□ The metric models cannot accept this command.

19

Advertising