ENMET Formaldemeter htV User Manual

Page 11

Advertising
background image

Formaldemeter htV

ENMET Corporation

9

5.5 Instrument Battery

With the exception of the battery, your

Formaldemeter

htV has no user-serviceable components. It is important that no

attempt is made to open the instrument other than to replace the battery. Any evidence of tampering with the instrument will
invalidate the warranty.

If you find that your instrument requires service or repair, please return it to the factory or an authorized PPM Service Center.

5.5.1 Low battery indicator

If

appears on the display when the instrument is switched on, then the battery voltage is too low and needs replacing

5.5.2 Battery Replacement

The instrument requires a 9V PP3 / MN1604 / 6LR61 type alkaline battery.

To replace the battery:

Remove the battery compartment cover at the bottom rear of the instrument by unscrewing the two screws.

Attach a new battery to the connector and place the battery in the compartment.

Secure the cover in place again with the two screws.

5.6 Cleaning the Instrument

Take care during cleaning
On no account should the

Formaldemeter

htV be immersed in liquid. Any fluid entering the instrument will destroy the

sensor and the electronic circuitry.

If the instrument’s enclosure requires cleaning, it should be wiped with a damp cloth. Never use abrasive or solvent based
cleaning agents.

5.7 Storage

When not in use, your

Formaldemeter

htV should be stored in the supplied carrying case in a clean, dry environment and

away from extremes of temperature.

5.8 Trouble Shooting

If display reads

and will not change after 3 minutes.

The gain/amplification of the signal form the sensor is so high that the signal never falls below a certain ‘threshold’ to enable it

to give the ready

display.

To remedy the problem a reset is needed on the instrument to erase the previous calibration.

This is achieved as follows:

Switch instrument

OFF

.

Hold both Cal buttons and at the same time switch the instrument

ON

.

The display will read

followed by

no

Press the Cal Up button to make the display read

yes

and then press the Sample button.

The display will then flash

[ ]

and then

and then show

Turn the instrument OFF and then back ON again to complete the Reset.

The instrument is no longer calibrated and will require recalibration.

Ensure that you follow the calibration procedure in section 5.4 carefully. Ensuring that the standard has equilibrated at a stable
temperature for at least 1 hour before calibrating and that the tube is not warmed excessively through handling.

Advertising