ENMET ISA-40 User Manual

Page 8

Advertising
background image

ENMET Corporation

ISA-40 Oxygen Monitor

6

3.2.2 Remote Sensor Assembly Wiring

See Figure 4 for location of terminal blocks and optional intrinsic safety barrier.
Sensor assemblies with intrinsic safety barriers cannot have an audio alarm.

Table 2: Remote Sensor Assembly Wiring (without Intrinsic Safety Barrier)

Channel

Signal

Control Unit Connection

Sensor Connection

1

Audio Alarm (–)

TB3 – 1

TBS – 1

1

Audio Alarm (+)

TB1 – 1 Horn

TBS – 2

1

Oxygen Cell (–)

TB1 – 1 O2 –

TBS – 3

1

Oxygen Cell (+)

TB1 – 1 O2 +

TBS – 4

2

Audio Alarm (–)

TB3 – 2

TBS – 1

2

Audio Alarm (+)

TB1 – 2 Horn

TBS – 2

2

Oxygen Cell (–)

TB1 – 2 O2 –

TBS – 3

2

Oxygen Cell (+)

TB1 – 2 O2 +

TBS – 4

3

Audio Alarm (–)

TB4 – 1

TBS – 1

3

Audio Alarm (+)

TB2 – 3 Horn

TBS – 2

3

Oxygen Cell (–)

TB2 – 3 O2 –

TBS – 3

3

Oxygen Cell (+)

TB2 – 3 O2 +

TBS – 4

4

Audio Alarm (–)

TB4 – 2

TBS – 1

4

Audio Alarm (+)

TB2 – 4 Horn

TBS – 2

4

Oxygen Cell (–)

TB2 – 4 O2 –

TBS – 3

4

Oxygen Cell (+)

TB2 – 4 O2 +

TBS – 4

Table 2A: Remote Sensor Assembly Wiring(with Intrinsic Safety Barrier)

Channel

Signal

Control Unit Connection

Sensor Connection

1

Oxygen Cell (–)

IS1 – 2

TBS – 3

1

Oxygen Cell (+)

IS1 – 3

TBS – 4

2

Oxygen Cell (–)

IS2 – 2

TBS – 3

2

Oxygen Cell (+)

IS2 – 3

TBS – 4

3

Oxygen Cell (–)

IS3 – 2

TBS – 3

3

Oxygen Cell (+)

IS3 – 3

TBS – 4

4

Oxygen Cell (–)

IS4 – 2

TBS – 3

4

Oxygen Cell (+)

IS4 – 3

TBS – 4

N

OTE

: These sensor assemblies cannot have an audio alarm.