ENMET ISA-44-2OD User Manual
Page 19

ISA-44-2-OD
ENMET Corporation
15
7. Put the cal cup of the calibration fixture over the sensor. Let the gas flow over the sensor until the alarm triggers.
Use Table 6 for the approximate allowable time period until alarm:
Table 7: Approximate Alarm Time
Gas or Vapor
Time to Alarm
20 ppm CO
5 - 7 minutes
50 ppm CO
5 minutes
200 ppm CO
2 minutes
10% LEL methane
1 minute
20% LEL methane
1 minute
100 ppm vinyl chloride
1 - 2 minutes
typical hydrocarbon vapors
1 - 2 minutes
20% LEL propane
1 minute
8. After you verify the alarm response, turn the gas off and remove the calibration fixture from the gas cylinder.
9. Turn the switch back to
OPERATE
.
C
AUTION
: If steps 1 - 10 do not trigger the alarms, and no electrical or mechanical malfunction is apparent, you must
recalibrate the instrument.