ENMET ISA-44-RALE-OD User Manual

Page 25

Advertising
background image

ISA-44RALE-OD/ISA-44RAHE-OD/ISA-44E-OD

ENMET Corporation

21

a. Place the Purge/Operate/Horn-off switch to PURGE. Measure purge voltage across TB4-7 (+) and TB4-6 (-)

inside the terminal enclosure. Voltage should be 1.65 V

DC

within a 0.03 V

DC

range. If necessary, adjust

potentiometer R38 (on CO channel circuit board, located on display plate) until the required voltage is obtained.
Place the Purge/Operate/Horn-off switch to OPERATE, typically the voltage should be .83 ± .03. Verify the
voltage listed on the label located on the instrument display plate. Return switch to PURGE and let run for 10
minutes.

b. When purging is complete, set the Purge/Operate/Horn-off switch to OPERATE and allow one half hour for the

MOS sensor to stabilize. Repeat steps 3 – 10. If unit still does not operate properly, the sensor may need to be
replaced.

6.2 Oxygen Channel Calibration

C

AUTION

: Enclosures are not explosion proof when the covers are removed. Check the surrounding atmosphere for

combustible gases and vapors before performing calibration procedures.

N

OTE

: This procedure is only necessary if the oxygen channel is not responding accurately or if the channel fails the

rough test.

N

OTE

: Check oxygen cell output, measure the voltage between the cell lead. This should be between 0.043 V

DC

and 0.09

V

DC

. If it is not, replace the sensor. See section 7.1 and perform a rough test see section 5.0.

Figure 8: Control Unit Oxygen Channel

PROCEDURE:
1. Set the Purge/Operate/Horn-off switch to HORN-OFF. Open the oxygen cell housing. Remove the cell from the

circuit board (do not touch the screen on the cell).

2. Connect the negative voltmeter lead to TP101 and the positive voltmeter lead to TP102 (on the oxygen channel

circuit board in the display plate).

3. Adjust the null adjust potentiometer R108 (on oxygen channel circuit board) so the voltmeter shows 0.00 V

DC

between test points TP101 and TP102.

4. LOW LEVEL SET ADJUST:

a) Reconnect the oxygen cell.

b) To check oxygen cell output, measure the voltage between the cell lead. This should be between 0.043 V

DC

and

0.09 V

DC

.

c) Connect the negative lead of the voltmeter to TP101 and the positive lead to TP103 in the main circuit board of

the monitor.

d) Adjust the low level set potentiometer, R112, to give .766 V

DC

between TP101 and TP103.

O

XYGEN

M

ETER

Alarm LED
Red

Power LED
Green

Horn Off LED
Red

R108

R110

R112

R116

TP101

TP102

TP103

TP101