0 installation, 2 mounting, 3 wiring – ENMET CD-1300-ST User Manual

Page 5

Advertising
background image

CO

2

Sensor / Transmitter

ENMET Corporation

3


3.0 Installation

Figure 1: CD-1300-ST Series, Mounting Dimensions

3.1 Removal of Cover

To open the

CD-1300-ST

series use a coin in the slot on the bottom to release the snap. Lift the cover up slightly to disengage

the closure and remove cover with a downward motion to clear the catch at the top of the unit. The locations of controls and
terminals on the main circuit board are shown in the Figure 2.

3.2 Mounting

The

CD-1300-ST

series is designed for flush mounting with two fasteners. The locations of the mounting points (shown in

Figure 1) allow direct mounting on a standard simplex (single circuit) junction box. There is a wiring cutout in the center of the
unit near the terminal strips.

3.3 Wiring

This section describes the external connections to the CD-1300-ST Series. Wiring enters the chassis through the cutout in the
center of the unit.

Figure 2: CD-1300-ST Series Component Location

3.200

2103

.

1.600

1.103

0.000

0

.0

0

0

0

.4

0

6

2

.0

0

3

2

.7

5

3

3

.5

6

2

5

.2

0

0

C

USTOMER

W

IRING

N

OTE

:

When connecting to

ENMET

CP-10, CP-60,

LC-8 or other 3-wire type controllers:

1.)

Preferred wiring for long runs,
use 4-wire cable, connect +24VDC, Power
Ground, Signal Ground & 4-20mA at
transmitter and Connect both Power
Ground & Signal Ground together to
ground at the Controller.

2.)

Alternative wiring for short runs,
use 3-wire cable and jumper Power Ground
and Signal Ground together inside the
transmitter.

Sensor

UP

SW1

SW3

SW2

DN

M

JP5
JP4
JP3
JP2
JP1

J5

J5

U3

Microprocessor

Upper Mounting
Screw Access

Wiring Access

+ 24VDC
Power Ground

Lower Mounting
Screw Access

Relay Contact

Signal Ground

4-20mA

Jumper Here for

Current O

UTPUT

Jumper Here for

Voltage O

UTPUT

Option Button

Down Button

Up Button

User Jumpers