Instrucciones de control remoto
Fuente de analizador de CD/PMD
91
Instrucciones específicas del producto: Descripción
SOURce:POWer[:STATe]
Description This command turns on or off the source. When the source is
on, the red LED on the front of the module illuminates.
Syntax
SOUR:POW[:STAT]<wsp><boolean>|ON|OFF
Parameters A boolean parameter:
“1”–turns the source on
“0”–turns the source off
Example
SOUR:POW:STAT 1 (turns the source on.)
SOURCE:POWER 0 (turns the source off.)
SOURce:POWer:STATe?
Description This query returns a value indicating the state of the optical
source (on or off).
Syntax
SOUR:POW:STAT?
Response
A boolean value:
“1”–the source is on
“0”–the source is off
Example
SOUR:POW:STAT? 1