EXFO FTB-3930 MultiTest User Manual

Page 85

Advertising
background image

Troubleshooting

Multifunction Loss Tester

79

Solving Common Problems

Unable to send a text

message.

Selected port differs

from actual port to

which fiber is

connected.

Receiving unit’s port is

in use.

Remote unit is not

compatible.

On the sending unit, set

the correct port (SM or

MM) in Setup (F

AS

T

ES

T

tab).

Wait until F

AS

T

ES

T

completes, turn source

and ORL meter off or exit

Probe mode (FOT-930).

Make sure remote unit is

an FOT-930, FTB-3930,

FOT-920 or FTB-3920.

On Results tab, a box

indicates “Invld” instead of

a numerical value.

The application is unable to

calculate the F

AS

T

ES

T ORL

value (e.g., at 1310 nm for

short fibers).

Perform the measurements

again.

On Results tab, a box

indicates “- - - -” instead of

a numerical value.

The calculated value makes

no sense (e.g., negative

ORL value).

Perform the measurements

again. Note that the

calculated value is usually

shown on the printed report.

When printing a report,

rightmost columns are not

printed.

The Fiber ID column is too

wide. The width of this

column is determined with
the widest name.

Rename fiber IDs that have a

long name.

When saving data in text

formats, some ORL values

are higher than the

indicated sensitivity.

Saved text files containing

symbols such as “>” may

not open properly in some

applications.

The actual measured value is

used instead, even if it is not

accurate.

Problem

Possible Cause

Solution