Configure[1..n]:analysis:threshold: sloss – EXFO FTB-7000 OTDR Series for FTB-500 User Manual
Page 415

Referencia de instrucciones SCPI
OTDR
401
Product-Specific Commands—Description
:CONFigure[1..n]:ANAlysis:THReshold:
SLOSs?
Description
This query returns the splice loss threshold that
will be used for the next acquisition.
*RST reverts this setting to default value.
Syntax
:CONFigure[1..n]:ANAlysis:THReshold:SLOSs?[<
wsp>MINimum|MAXimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<Splice Loss>
Response(s)
Splice Loss:
The response data syntax for <Splice Loss> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the splice loss threshold.
Example(s)
CONF:ANA:THR:SLOS 0.03
CONF:ANA:THR:SLOS? Returns 0.03