SCPI Command Reference
52
FTB-9100
Product-Specific Commands—Description
:ROUTe[1..n]:SCAN:ADJust:AUTO?
Description
This query returns the current value of the
<Optimize for repeatability> option. When this
option is selected, the switch mechanism returns
to a reference position before aligning to a new
position. This ensures optimum repeatability.
When this option is not selected, the switch
mechanism goes directly to the new channel
position. This provides faster switching times, but
less repeatability.
On *RST, the value of this setting is OFF.
Syntax
:ROUTe[1..n]:SCAN:ADJust:AUTO?
Parameter(s)
None
Response Syntax
<AutoAdjust>
Response(s)
AutoAdjust:
The response data syntax for <AutoAdjust> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.