Hanna Instruments HI 98170 User Manual

Page 33

Advertising
background image

3 3

CHR xx

Change the instrument range according with the parameter
value (xx):

• xx=01 pH range/0.01 resolution
• xx=02 pH range/0.1 resolution (HI 98170, HI 98171)
• xx=03 mV range
• xx=04 Relative mV range
• xx=05 ISE range (HI 98172)

The instrument will answer for these commands with:

<STX> <answer

> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER

The instrument will answer for these commands with:

<STX> <answer> <checksum> <ETX>

where the checksum is the bytes sum of the answer string sent as 2 ASCII characters.
All the answer messages are with ASCII characters.
RAS

Causes the instrument to send a complete set of readings in
according with the current range:

• pH and temperature reading (HI 98170), and mV

reading (HI 98171) on pH range.

• Rel mV, absolute mV and temperature reading on Rel

mV range (HI 98171, HI 98172).

• ppm, mV and temperature reading on ppm range

(HI 98172).

The answer string contains:

• Meter mode (2 chars):

• 01 - pH range (0.01 resolution)
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range
• 05 - ISE range

• Meter status (2 chars of status byte): represents a 8 bit

hexadecimal encoding.