CHR xx

Change the instrument range according with the parameter
value (xx):

• xx=01 pH range/0.01 resolution
• xx=02 pH range/0.1 resolution
• xx=03 mV range
• xx=04 Relative mV range

The instrument will answer for these commands with:

<STX> <answer

> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:

<STX> <answer> <checksum> <ETX>

where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS

Causes the instrument to send a complete set of readings in
according with the current range:

• pH and temperature reading (HI 98140), and mV

reading (HI 98150 & HI 98160) on pH range.

• Rel mV, absolute mV and temperature reading on

Rel mV range (HI 98150 & HI 98160).

The answer string contains:

• Meter mode (2 chars):

• 01 - pH range (0.01 resolution)
• 02 - pH range (0.1 resolution)
• 03 - mV range
• 04 - Rel mV range

• Meter status (2 chars of status byte): represents a 8

bit hexadecimal encoding.
• 0x10 - temperature probe is connected
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter