Hi 4003 cadmium half-cell – Hanna Instruments HI 4003 User Manual

Page 2

Advertising
background image

2

HI 4003 Cadmium Half-cell

HI 4003 Cadmium Half-cell

HI 4003 Cadmium Half-cell

HI 4003 Cadmium Half-cell

HI 4003 Cadmium Half-cell
HI 4103 Cadmium Combination Electrode

HI 4103 Cadmium Combination Electrode

HI 4103 Cadmium Combination Electrode

HI 4103 Cadmium Combination Electrode

HI 4103 Cadmium Combination Electrode

I.

I.

I.

I.

I. Introduction:

Introduction:

Introduction:

Introduction:

Introduction:

The Hanna HI 4003 and HI 4103 are ion selective elec-
trodes designed for the measurement of cadmium ions in
aqueous solutions. The HI 4003 is a solid state half-cell
sensor that requires a separate reference. The HI 4103 is a
combination ion selective electrode.

II

IIII

IIII.

Specifications

Specifications

Specifications

Specifications

Specifications

Type:

Solid State electrode with
a Cadmium sulfide/ silver
sulfide membrane.

Ion(s) measured:

Cadmium (Cd

2+

+

+

+

+

)

Measurement range: 0.1

M to 1X 10

-7

M

11,200 to 0.01 ppm

Interfering ions:

Silver and mercury must

be absent. Any ion that forms a more insoluble sulfide than
cadmium sulfide will interfere with the electrode response.
These include lead (Pb

2+

)and cupric ions (Cu

2+

). Oxidation

of the cadmium surface by ferric ion( Fe

3+

) will severely

poison the electrode. Ions that can form complexes with
cadmium ions will reduce the measured concentration.
Examples include bromide, chloride and EDTA.

Operating Temperature:

0-80

°C

Operating pH:

2 to12 pH

concentration dependant

(4.5-7.5 recommended)

Dimensions:

12 mm (OD) X 120 mm
nominal insertion
(0.47

” X 4.72”)

Connection:

BNC

Advertising