Hanna Instruments HI 4101 User Manual

Page 35

Advertising
background image

15

Typical calibration curve for HI 4101 Ammonia ISE

•

Standards should bracket the measurement range of
interest and differ from each other by a factor of 10 in
the linear regions.

•

Standards and solutions should be at the same tem-
perature. 2 mL of ISA is added to each 100 mL of
sample and standard. A color change should occur to
indicate the ISA has been added.

•

Protect these solutions from loss of dissolved gas by
covering and using promptly.

3) Follow section VII; General Guidelines to optimize test

set-up.

4) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable reading before
reading/recording values.

5) To prevent carry over and contamination of samples,

rinse sensors with deionized water and blot dry be-
tween samples.

6) Between measurements suspend sensor tip in a small

sample of NH

3

Conditioning solution;

HI 4001-45. Rinse body with deionized water and

blot dry before placing in next sample.

Typical Linearity for HI 4101 Electrode

-300

-250

-200

-150

-100

-50

0

50

0

1

2

3

4

5

6