Change the instrument range according with the parameter
value (xx):

• xx=10 pH range/0.001 resolution on channel 1
• xx=11 pH range/0.01 resolution on channel 1
• xx=12 pH range/0.1 resolution on channel 1
• xx=13 mV range on channel 1
• xx=14 Relative mV range on channel 1
• xx=15 ISE range on channel 1 (HI 3221)
• xx=20 ISE range on channel 2 (HI 3222)
• xx=21 mV range on channel 2 (HI 3222)
• xx=22 Relative mV range on channel 2 (HI 3222)

The instrument will answer for these commands with:

<STX> <answer

> <ETX>

where:

<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)

COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:

<STX> <answer> <checksum> <ETX>

where the checksum is the bytes sum of the answer string sent as 2 ASCII characters.
All the answer messages are with ASCII characters.
RAS

Causes the instrument to send a complete set of readings in
according with the current range:

• pH, temperature and mV reading on pH range.
• Rel mV, absolute mV and temperature reading on Rel

mV range.

• concentration, mV and temperature reading on ppm

range (HI 3221, HI 3222).

The answer string contains:

• Meter mode (2 chars):

• 10 pH range/0.001 resolution on channel 1
• 11 pH range/0.01 resolution on channel 1
• 12 pH range/0.1 resolution on channel 1