Hanna Instruments HI 253 User Manual

Page 16

Advertising
background image

16

• The pH calibration slope (the GLP slope is the average of the calibration

slopes; the percentage is referred to the ideal value of 59.16 mV/pH).

• The pH calibration buffers in calibrating order.

The first pH calibration buffer:

The second pH calibration buffer:

The third pH calibration buffer:

Note:

•

The “OLd” message displayed beside the pH value means that

this buffer was not used during last calibration. Press and hold
down SET/CLR if you want to see calibration date (or time if old
calibration was made in the same day with current calibration).

•

If “no bUF” message appears on the LCD, the instruments inform

you that calibration was performed in less than three points.

• pH resolution during calibration.

• Calibration Alarm Time-Out status:

if disabled.

CAL

pH

BUF

CAL

pH

BUF

%

CAL

CAL

BUF

CAL

CAL

pH

BUF