Hanna Instruments HI 2216 User Manual
Page 35

35
PC INTERFACE
PC INTERFACE
PC INTERFACE
PC INTERFACE
PC INTERFACE
Data transmission from the instrument to the PC can be done with the
HI 92000 Windows
®
compatible software (optional).
HI 92000 also
offers graphing and an on-line help feature.
Data can be exported to the most popular spreadsheet programs for
further analysis.
To connect your instrument to a PC, use a standard USB cable connector.
Make sure that your instrument is switched off and plug one connector to
the instrument’s USB socket and the other to the USB port of your PC.
Note: If you are not using Hanna Instruments HI 92000 software,
please see the following instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal
program. Use a standard USB cable to connect the instrument to a PC,
start the terminal program and set the communication options as follows:
8, N, 1, no flow control, 9600 baud rate.
COMMAND TYPES
To send a command to the instrument follow the next scheme:
<command prefix> <command
> <CR>
where:
<command prefix> is a 16 ASCII character
<command> is the command code.
Note: Either small or capital letters can be used.
SIMPLE COMMANDS
RNG
Is equivalent to pressing
RANGE
CAL
Is equivalent to pressing
CAL
CFM
Is equivalent to pressing
CFM
UPC
Is equivalent to pressing the
UP arrow key
DWC
Is equivalent to pressing the
DOWN arrow key
LOG
Is equivalent to pressing
LOG
RCL
Is equivalent to pressing
RCL
SET
Is equivalent to pressing
SETUP
Notes:• If the reading is not close to the selected calibration point,
“
WRONG” tag will blink. Verify calibration condition or
contact your vendor if you can not calibrate.
• Pressing
CAL key during temperature or mV calibration the
instrument quit calibration mode and restart to measurement
mode without memorizing calibration.