• Select the appropriate measurement range (EC or TDS).

• Wait for a couple of minutes for the

temperature sensor to reach thermal
equilibrium. The display then shows the
measurement automatically compensated
for temperature with the appropriate
following indication: “

mS” tag indicates

the meter is in EC mode, while “

ppm”

tag indicates the meter is in TDS
mode.

TAKING TEMPERATURE MEASUREMENTS

• Submerge the tip (4cm/1½") of the

probe into the sample to be tested.

• Select the °C mode.

• Stir briefly and wait a couple of minutes

for the reading to adjust and stabilize.
The display shows the temperature
value.

Notes: • If the display shows only a “1”

on the far left hand side, the
reading is out of range.

• It is recommended to clean

often the probe with

HI 700661

Cleaning Solution.

• After measurements have been completed, the instrument

should be switched off, and the probe cleaned and
covered with the protective cap.