Hanna Instruments HI 93703 User Manual

Page 6

Advertising
background image

11

10

HOW TO ENSURE ACCURATE MEASURE-
MENTS

• Each time the cuvette is used, tighten the

cap to the same degree.

• Discard the sample soon after the reading is

taken to avoid permanently clouding the glass.

• All glassware used to contain the standards

and the samples should be maintained clean,
washed with HI 93703-50 cleaning solution
and rinsed with HI 93703-0 or turbidity-free
water.

• Collect the samples in clean glass or plastic

bottles, fit stoppers and perform the analysis
quickly. If unavoidable, store the sample in
a cool, dark place, but not for longer than 24
hours (the sample needs to be kept at room
temperature prior to the analysis).

• To obtain a representative sample, gently,

but thoroughly, mix it before samples are
taken. Do not shake (to prevent air bubbles)
and do not let the sample settle.

• It is recommended to monthly calibrate the

meter with the supplied HI 93703-10 @10
FTU* standard or more frequently for greatest
accuracy.

• Before inserting vials

into the instrument,
wipe them with HI
93703-70
soft, lint-free
tissue. Handle vials so
that no fingerprints can
get on the areas
where light passes (ap-
proximately 2 cm/1
inch from the bottom
of the vial).

If you experience any problems in taking mea-
surements, please contact your dealer or the
nearest Hanna Instruments Customer Service
Department.

* 1 FTU = 1 NTU

* 1 FTU = 1 NTU

• Even though HI93703 covers a very wide

range of turbidity values, for very accurate
measurements of samples exceeding
40 FTU *, Standard Methods require dilu-
tion. In such cases, the correct amount of
HI 93703-0 or turbidity-free water to be
added to the sample can be calculated
as follows:

Vos = 3000 / T

where: Vos = volume of sample (mL) to be

combined with HI 93703-0 to
obtain the final volume of 100
mL.

T

= HI 93703 reading (exceeding

40FTU*)

E.g.: HI 93703 reading = 200 FTU *

3000 / 200 = 15 mL (Vos)
15 mL (Vos)+85 mL (HI 93703-0) = 100

mL

At this point, take a sample of this solution
and measure turbidity.
The correct turbidity value of the original
sample will be:

T

n

x 100 mL / Vos = T

a

where: T

n

= new HI 93703 reading

T

a

= actual turbidity value of the original

sample

Advertising