Err1, Operational guide – Hanna Instruments HI 93703 User Manual

Page 8

Advertising
background image

8

To prepare the instrument for taking
measurements, first install the batteries (see
Battery Replacement section on page 20) and
then turn the instrument on.

To maximize battery-life, the display is
automatically switched off after 5-minutes of
non-use. To reactivate the display, simply
press the ON/OFF key.

MEASUREMENT PROCEDURE:

• Turn the meter on by

pressing the ON/OFF
key.

• The meter will carry

out a self-test display-
ing a full set of figures.

• The meter will carry

out a battery test show-
ing in percentage the
battery life left.

• When the LCD dis-

plays “----” the meter
is ready to measure.

• Fill a clean cuvet up to one quar-

ter inch (0.5 cm) from its rim with
the thoroughly agitated sample.

• Allow sufficient time for bubbles

to escape before securing the cap.
Note: do not overtighten the cap.

LOGGING

MICROPROCESSOR

TURBIDITY METER

LOGGING

MICROPROCESSOR

TURBIDITY METER

HI 93703-11

VIEW

VIEW

STO

STO

CLR

CLR

GLP

GLP

CAL

CAL

DATE

DATE

READ

READ

ALT

ALT

OFF

ON

1/4“ 0.5

cm

%

ba

98

OPERATIONAL GUIDE

LOG

LOBAT

TIME

DATE %

188

17

• If "ERR1" appearson

the LCD, please check
the standard solution.

• After approximately 30 seconds the meter

will display 10.00, prompting the user to
place the 10.00 FTU standard solution in
the cuvet holder.

• Place the 10.00 FTU

standard in the holder press CAL, SIP and
CL will start blinking

• After approximately

30 seconds the meter
will display 500, asking the user to place
the 500 FTU buffer
solution in the cuvet
holder.

NOTE: At this point the

user can save the two point calibration
setup by pressing ALT & CAL buttons,
leaving the calibration mode.

To perform a three point
calibration, place the 500
FTU standard solution in the
cuvet holder.

• Press CAL, SIP and

CL will start blinking

10.00

cl

sip

cl

500

cl

sip

cl

err1